Skip to content

Follow the metaboigniter workflow

The primary workflow analyzes the biological, pooled QC, and blank injections together so their features share one correspondence rule. The standard mixtures remain separate. Sparse known mixtures should not determine how complex biological extracts align.

The selected workflow is nf-core/metaboigniter 2.0.1, fixed at commit 55d82547604fcae3b6557fe7a3c442b623184f34. The execution record reports acceptance and resource evidence separately from this explanation. It remains the authority for which complete run supplies the downstream fixtures.

The source assay supplies negative centroid MS1 mzML files. The preparation checks complete byte counts and provider SHA-256 values, then reads every scan and binary array. Independent mzR and pyOpenMS checks compare scan counts, retention times, and decoded peaks under declared tolerances.

The pipeline receives 30 unique file aliases. Each alias has the same bytes as its original deposited file, and a manifest preserves the original filename, source hash, and analytical role. This matters because the selected release derives its internal identifier from the input filename rather than the samplesheet’s sample label.

The source acquisition includes calibration near 18 minutes. The preparation preserves those scans. The feature review records its biological retention-time interval and any exclusions rather than silently changing the deposited source.

Stage Input and output Required review
Feature detection FeatureFinderMetabo converts compatible mass traces into featureXML records. Inspect mass tolerance, chromatographic width, integrated area, and weak or split signals.
Retention alignment MapAlignerPoseClustering estimates transformations between feature maps. Inspect the reference choice, matched landmarks, transformation residuals, and QC overlays.
Spectrum transformation MapRTTransformer applies the declared retention transformations. Confirm that transformed spectra and feature coordinates use the same time scale.
Adduct annotation MetaboliteAdductDecharger evaluates candidate ion relationships. Preserve the candidate relationships without converting them into compound identities.
Feature correspondence FeatureLinkerUnlabeledKD connects compatible observations across inputs. Check map identities, duplicate assignments, mass differences, and retention differences.
Table export The workflow exports the linked consensus as a quantification table. Verify quantities against the underlying consensus handles and preserve missing observations.

The release output guide describes these distinct products. A flat table cannot replace their provenance. In particular, an absent feature observation is not a measured zero concentration.

The workflow explicitly sets negative polarity and skips another centroid step. Mass-trace quantification uses integrated area. It preserves convex hulls and intermediate files for review.

Requantification remains disabled in the first analysis. The feature table therefore exposes non-detections before any targeted attempt to recover an omitted signal. An additional extraction method would need its own parameters and evidence.

SIRIUS and MS2Query remain disabled because the selected primary cohort contains MS1 measurements. The repository’s separate pooled MS/MS files do not enter this run. Their existence cannot establish that this analysis obtained fragment evidence for an unknown feature.

The local preflight uses two plants, two pooled QC injections, and one blank from the selected dataset. It exercises the negative-mode path with the same parameter file and fixed container manifests used by the complete cohort.

Execution acceptance requires successful fresh processes, complete feature maps, correct consensus sample identities, and preserved commands. Scientific acceptance then examines background, repeatability, correspondence, and missing observations. A process exit code alone cannot establish those properties.

Continue with feature detection to inspect the shared measurements, then quality filters to define which features support the species comparison.

  • nf-core/metaboigniter usage at the compared revision

    The guide defines sample, level, type, and msfile columns and describes MS1 quantification. Its alignment and optional requantification sections explain which processing decisions precede the feature table used by later lessons.