Affinity Platforms Compared
One idea, four readouts
Section titled “One idea, four readouts”Every affinity platform rests on one idea. A binder recognizes a protein, and a second step turns that binding into a signal an instrument can read. The platforms differ in the binder, in how many proteins they read from one sample, and in the signal. Those three choices decide what lands in the exported file and which problems the analysis inherits.
The enzyme immunoassay was published twice in 1971, independently, by Engvall and Perlmann at Stockholm University, who called it ELISA, and by Van Weemen and Schuurs in the Netherlands. Lequin 2005 tells the story. The principle has not changed since. Something is fixed to the plastic of a well, the sample binds to it, unbound material is washed away, and an enzyme carried by an antibody turns a substrate into a colored product.
The formats differ in what sits on the plate and what carries the enzyme. Thermo Fisher’s overview of ELISA describes all four.
| Format | On the plate | Carries the enzyme | Signal as the protein rises |
|---|---|---|---|
| Direct | the antigen | the primary antibody | rises |
| Indirect | the antigen | a secondary antibody that binds the primary | rises |
| Sandwich | a capture antibody | a detection antibody against a second epitope | rises |
| Competitive | a capture antibody | labeled antigen that competes with the sample’s antigen | falls |
The same overview names the sandwich as the most widely used format. Two antibodies must bind two different epitopes of the same molecule, and the overview credits the format’s popularity to its sensitivity and specificity. The competitive format suits small antigens with only one epitope, where two antibodies cannot bind at once.
well surface ── capture antibody ── protein ── detection antibody ── enzyme │ substrate ──► colored productA common pairing is horseradish peroxidase with the substrate TMB, which turns blue. An acid stop solution turns it yellow and fixes the end point, and the plate reader measures absorbance. An R&D Systems DuoSet protocol reads each well at 450 nm and subtracts a reading at 540 or 570 nm, which corrects for optical imperfections in the plate, and warns that uncorrected readings may be higher and less accurate. The export holds an optical density for each well, sometimes at both wavelengths.
The concentration comes from standards on the same plate. The DuoSet protocol recommends a seven point series of twofold dilutions, run in duplicate, and a four parameter logistic fit. The ELISA standard curve page fits that curve and reads samples off it.
Luminex xMAP
Section titled “Luminex xMAP”Luminex measures many proteins in one well by running a separate sandwich assay on each of several sets of beads. The approach goes back to FlowMetrix, a Luminex Corporation system that Fulton et al. 1997 described, which read 64 sets of fluorescent microspheres on a benchtop flow cytometer.
Each bead is dyed inside with a set ratio of fluorescent dyes, and that ratio is its region. According to Luminex’s own description, MicroPlex beads are 5.6 µm across and carry two dyes, for about 100 regions. MagPlex beads are 6.5 µm, superparamagnetic, and carry up to three dyes, for 500 regions. One capture antibody is coupled to each region, the sample binds, and a biotinylated detection antibody follows. Streptavidin conjugated to phycoerythrin, PE, is the reporter, as Houser 2012 describes.
The instrument reads two colors from every bead. A red laser or LED excites the internal dyes and identifies the region, and a green laser or LED excites the reporter, according to the xMAP technology page. In a flow reader the beads pass in single file through a flow cell, and Houser 2012 gives a 635 nm classification laser and a 532 nm reporter laser for Bio-Rad’s flow based readers. A MAGPIX works differently. It holds magnetic beads in place with a magnet, lights them with LEDs, and images them with a CCD camera.
The reader also sets the plex. The MagPlex product page gives up to 50 analytes in one well for the MAGPIX and up to 100 for the Luminex 200, and only the FLEXMAP 3D and the xMAP INTELLIFLEX read all 500 regions.
For each well and each analyte, the software reports how many beads it read and the median reporter fluorescence across them, the MFI. Concentrations come from a standard curve for each analyte, which the acquisition software fits as a five parameter logistic. The Luminex format page shows the xPONENT export column by column and the bead count check that comes before anything else.
Olink proximity extension assay
Section titled “Olink proximity extension assay”The proximity extension assay, PEA, replaces the enzyme with DNA. Lundberg et al. 2011 link each antibody of a matched pair to a DNA strand. When both antibodies bind the same protein, the two strands come close enough to hybridize, and a DNA polymerase extends them into a new sequence that PCR can amplify and count.
antibody A + oligo a ──┐ ├──► both bind one protein ──► oligos hybridizeantibody B + oligo b ──┘ │ polymerase extends them into a reporter │ qPCR on Target · sequencing on ExploreA reporter forms only when the two strands meet, so the specific signal needs both antibodies on one molecule. Strands that meet by chance give a background, which negative controls on every plate measure. Assarsson et al. 2014 argue that the need for two reporters protects against antibodies that bind the wrong protein, the problem that limits other multiplex immunoassays. Their 96-plex design used 1 µL of sample. The current Olink Target 96 panels measure 92 proteins from 1 µL and report NPX in relative units.
On Target panels the readout is qPCR. Olink’s normalization white paper derives NPX in three steps. The Ct of an extension control in the same well is subtracted from the Ct of each assay, the result is set against an inter plate control, and a correction factor fixed at validation inverts the scale. A higher NPX then means more protein, on a log2 scale where one unit is roughly a doubling.
Olink Explore swaps qPCR for sequencing. Wik et al. 2021 describe nearly 1,500 assays in four 384-plex panels, read on an Illumina NovaSeq 6000 for 96 samples at a time. Samples are diluted into blocks of proteins with similar abundance, and the whole protocol uses 2.8 µL of plasma. NPX is computed from read counts by normalizing to the extension control, taking the log2, and adjusting with a plate control of pooled plasma. Olink Explore HT extends this to about 5,400 proteins from 2 µL.
Every Olink assay carries its own limit of detection. Negative controls of buffer on each plate set the background, and the OlinkAnalyze vignette defines LOD as three standard deviations above it. For Explore data, Olink’s LOD tutorial takes the median of the negative controls and adds three standard deviations or 0.2 NPX, whichever is larger. Values below LOD are still reported, and the same tutorial recommends using all the data downstream. The Olink NPX format page shows where LOD and the QC flag sit in the export.
SomaScan, for comparison
Section titled “SomaScan, for comparison”SomaScan uses no antibodies. Its binders are SOMAmers, slow off rate modified aptamers, short DNA strands with chemically modified nucleotides that fold into shapes that bind proteins. Gold et al. 2010 introduced the assay, which turns a pattern of protein concentrations into a pattern of aptamer concentrations and reads that pattern on a DNA microarray. Their version measured 813 proteins from 15 µL of serum or plasma.
The current SomaScan 11K technical note lists about 11,000 protein measurements covering more than 10,000 human proteins, read across three serial dilutions of each sample. Its output, RFU, is proportional to the amount of target epitope, which makes it relative in the same sense as NPX. SomaScan comes in a later release of this section.
The platforms side by side
Section titled “The platforms side by side”| Platform | Plex in one sample | Sample volume | Readout | Output unit | Absolute or relative | Typical use |
|---|---|---|---|---|---|---|
| ELISA | 1 | 100 µL of sample or standard per well in a DuoSet protocol | absorbance at 450 nm | concentration, such as pg/mL | absolute against the kit calibrator | one protein, measured well, across many samples |
| Luminex | up to 50 on MAGPIX, 100 on Luminex 200, 500 on FLEXMAP 3D | 12.5 µL for the Bio-Plex Pro human cytokine panels | PE fluorescence on each bead | MFI, then a concentration | absolute against the kit calibrator | cytokine and chemokine panels |
| Olink Target 96 | 92 | 1 µL | qPCR | NPX, log2 | relative | a focused panel, such as inflammation |
| Olink Explore | about 1,500 on Explore 1536, about 5,400 on Explore HT | 2.8 µL on Explore 1536, 2 µL on Explore HT | sequencing counts | NPX, log2 | relative | discovery in large cohorts |
| SomaScan | 813 in 2010, about 11,000 on the 11K assay | 15 µL in 2010, 130 µL listed for the 11K assay | fluorescence on a DNA microarray | RFU | relative | discovery in large cohorts |
The Luminex volume is the figure Bio-Rad gives for its Bio-Plex Pro human cytokine panels. Houser 2012 puts the cost of the alternative at more than 1 mL of serum or plasma to measure 27 cytokines by ELISA at 50 µL a well. The SomaScan 11K volume is the per sample figure in Table 1 of its technical note.
Start from what you need. If you need a concentration you can put next to a reference range, only ELISA and Luminex give one, and only against their calibrators. If you need thousands of proteins from a few microliters, Olink is the platform in the table built for it, SomaScan reaches further at a larger volume, and both give relative values in return. Luminex sits between them, with tens of proteins and a standard curve for each.
Next steps
Section titled “Next steps”The next page, Experimental design, covers the plate layout, controls and sample handling that every one of these platforms depends on.
Further reading
Section titled “Further reading”The OlinkAnalyze guide explains NPX, quality flags, and normalization. Read these definitions alongside the platform table to distinguish a relative assay signal from a calibrated concentration.
The Olink limit of detection tutorial explains how negative controls define the detection limit and how to handle values below it. It provides the detail behind the Olink readout described above.