What metabolomics measures
Untargeted LC-MS records ion signals that change over chromatographic time. A detected feature describes a measured mass-to-charge ratio, a retention time, and an intensity. That description does not identify a compound.
One compound can produce several features through isotopes, adducts, fragments, and different charge states. Different compounds can also share a mass or overlap chromatographically. The analysis must preserve these ambiguities when it moves from a spectrum to a statistical result.
Follow the measurement
Section titled “Follow the measurement”Liquid chromatography separates an extract before ions enter the mass spectrometer. In an MS1 spectrum, the instrument measures precursor ion signals. Repeated spectra trace those signals across an elution peak. A feature detector connects compatible mass traces and estimates their areas.
Retention time varies between injections, so the workflow aligns runs before it links corresponding features. A successful numerical match still needs quality checks. A wide matching window can connect unrelated peaks, while a narrow window can split one reproducible signal into several incomplete features.
The MTBLS865 example starts from deposited centroid mzML files measured in negative ion mode. Centroid data retain estimated peak positions and intensities. They do not retain the complete profile signal from which the instrument or conversion software estimated those peaks.
Separate targeted and untargeted questions
Section titled “Separate targeted and untargeted questions”| Approach | Declared measurement | Required evidence |
|---|---|---|
| Targeted quantification | The assay measures a specified compound or transition. | Calibration, selectivity, recovery, and suitable standards establish the quantitative claim. |
| Untargeted feature analysis | The workflow measures reproducible ion features within its detection and matching rules. | Blanks, pooled QC, mass accuracy, retention time, and detection patterns support the feature table. |
| Compound identification | The analysis assigns a molecular identity to a feature. | Authentic standards, retention behavior, and diagnostic spectra establish the supported resolution. |
The teaching example compares relative feature areas between two plant species. It has neither a calibration series for every feature nor verified identities for the biological feature universe. Its results therefore concern measured features, not absolute concentrations or a complete catalog of metabolites.
Keep the units attached to the result
Section titled “Keep the units attached to the result”Integrated ion signal depends on extraction, injection volume, ionization, detector response, and the sample itself. Two compounds at the same concentration can produce different signals. A matrix effect can change a compound’s signal without changing its amount.
Normalization adjusts a declared reference scale. It cannot recover an unknown extraction yield or turn a detector response into molar concentration. In this example, normalization and drift retain the original areas and compare several defensible scales.
The field study collected root exudates from plants. A difference between the selected species can reflect plant size, collection responses, and chemical composition. Without a verified mass or flux denominator, the tutorial does not interpret an intensity ratio as a ratio of exudation rates.
Treat identification as a separate question
Section titled “Treat identification as a separate question”An accurate mass can suggest molecular formulas, but several structures can share a formula. An adduct relationship supports an ion explanation without proving a structure. A small statistical p-value adds evidence about a feature difference, not its identity.
The identification lesson preserves the evidence behind each annotation. The interpretation lesson keeps unknown features visible and explains why they cannot support a named pathway result.
Scientific references
Section titled “Scientific references”Dietz et al. 2020 describe the source root-exudate study under CC BY 4.0. The original MTBLS865 protocols remain separate from the selected teaching analysis and its software.
Further reading
Section titled “Further reading”-
The original deposit describes the ionization mode, standard mixture, pooled controls, and blanks for this study. It connects the general measurement concepts on this page to a documented experiment.
-
nf-core/metaboigniter usage at the compared revision
The workflow documentation follows mass-spectrometry inputs through feature alignment and quantification. Use it to understand how a feature matrix arises before interpreting a feature as a named metabolite.