What is Proteomics
What protein measurement adds
Section titled “What protein measurement adds”RNA-seq counts transcripts, and a transcript is an instruction. The protein it codes for is made, modified, cut, secreted and broken down on schedules of its own, so a transcript count is an indirect measure of the protein. Schwanhäusser et al. 2011 measured mRNA and protein abundance and turnover for more than 5,000 genes in mammalian cells and found that protein abundance is controlled mainly at the level of translation. A later review by Liu, Beyer and Aebersold 2016 concluded that in many situations transcript levels alone are not enough to predict protein levels, especially while a cell is changing state.
Blood widens the gap. A plasma sample holds proteins that tissues all over the body secreted or leaked, and Anderson and Anderson 2002 describe plasma as containing all tissue proteins as leakage markers, next to the classical plasma proteins and a vast number of distinct immunoglobulins. A cytokine in plasma reports on cells that are not in the tube. There is no matching transcript to sequence, and the protein is all you can measure. Blood is also the sample medicine looks at first, and Geyer et al. 2017 call the clinical analysis of blood the most widespread diagnostic procedure in medicine.
Two ways to measure proteins
Section titled “Two ways to measure proteins”The methods split into two families, and the split decides which proteins you can see and what units you get back.
affinity assay sample ──► a binder recognizes ──► a signal for each ──► one number for each chosen protein target each protein
mass spectrometry sample ──► proteins digested ──► LC-MS/MS ──► peptides ──► proteins into peptides spectra identified inferredA targeted affinity assay uses a binder that recognizes one protein. The binder is an antibody in ELISA, Luminex and Olink, and a chemically modified DNA aptamer in SomaScan. You choose the list of proteins before the run, the assay reports only that list, and its sensitivity comes from how tightly and how cleanly the binder holds its target. The Affinity platforms page describes how each platform turns binding into a signal.
Discovery mass spectrometry needs no binder. Proteins are digested into peptides, the peptides are separated by liquid chromatography and ionized by electrospray, and the instrument records their masses and the masses of their fragments. Software then identifies the peptides by searching a sequence database, as Geyer et al. 2017 describe. Since a peptide’s mass and sequence identify it, the authors call the method inherently specific, and it can find proteins nobody planned to measure. The mass spectrometry files page covers the raw formats.
What each family measures and reports
Section titled “What each family measures and reports”Each platform records a physical signal and reports a number derived from it. The two are not the same, and the gap between them is where most of the analysis on the later pages happens.
| Family | What the instrument records | What you analyze | Scale |
|---|---|---|---|
| ELISA | absorbance of each well at 450 nm, the optical density | a concentration such as pg/mL, read off a standard curve | linear, absolute against the kit calibrator |
| Luminex | reporter fluorescence on each bead, summarized as the median fluorescence intensity, MFI, for each bead region | a concentration for each analyte, read off its own standard curve | linear, absolute against the kit calibrator |
| Olink | qPCR cycle thresholds on Target panels, sequencing counts on Explore | NPX, Normalized Protein eXpression | log2, relative |
| SomaScan | fluorescence of aptamers hybridized to a DNA microarray | RFU, relative fluorescent units | linear, relative |
| Mass spectrometry | mass spectra of peptides and their fragments | an intensity for each peptide or protein | linear, relative unless spiked standards are added |
ELISA and Luminex follow the same logic. Every plate carries a dilution series of a standard with known concentrations, a curve is fitted through the standards, and each sample’s signal is read back off the curve. Olink and SomaScan skip the curve. Olink normalizes its signal against controls in every well and on every plate, and SomaScan reports a fluorescence that its vendor states is directly proportional to the amount of target epitope in the sample. Mass spectrometry reports peptide intensities. Label free methods such as MaxLFQ compare them across runs, and tandem mass tags compare samples within one run through reporter ions.
Relative and absolute quantification
Section titled “Relative and absolute quantification”An absolute result is a concentration in physical units. A relative result tells you how one sample compares with another for the same protein, and nothing more.
ELISA and Luminex give concentrations, but absolute here means absolute against the kit’s calibrator. An R&D Systems DuoSet ELISA, for example, is calibrated against a purified recombinant protein that the company makes. The endogenous protein in your sample can differ from that recombinant standard in its form and its binding, so two kits for the same cytokine need not report the same value for the same sample.
Olink NPX is relative by design. Olink’s normalization white paper states that two proteins with the same NPX can have different absolute concentrations, that NPX should be compared for one assay at a time between samples, and that NPX from different runs should not be compared without inter plate normalization. One NPX unit is roughly a doubling of the protein, whatever the protein. SomaScan RFU are relative in the same sense, and its technical note describes the assay as quantifying the relative levels of proteins.
Mass spectrometry is relative unless you add a reference. The AQUA method spikes synthetic peptides labeled with stable isotopes into the digest and measures the native peptide against them, and Geyer et al. 2017 note that absolute quantification generally requires such heavy internal standards.
Most biological questions only need relative values. A difference between groups, a change over time, or a correlation with an outcome all work within one assay. What relative values cannot do is tell you whether protein A is more abundant than protein B, or put your numbers next to a clinical reference range.
The dynamic range of plasma
Section titled “The dynamic range of plasma”Plasma is hard to measure because its proteins span an enormous range of concentrations. Anderson and Anderson 2002 put more than 10 orders of magnitude between albumin and the rarest proteins measured clinically. Geyer et al. 2017 give the two ends. Albumin is about 50 mg/ml and roughly half of the total protein mass, while baseline IL-6 is about 5 pg/ml.
This range hurts mass spectrometry most. In data dependent acquisition the instrument fragments peptides in order of intensity, so the abundant proteins claim most of the spectra, and Geyer et al. 2017 report that the range still makes it difficult to identify more than a few hundred of the most abundant plasma proteins by LC-MS/MS. Studies often deplete the top 1 to 20 proteins first, on columns of antibodies against them.
Affinity assays meet the same range in another form. Each assay has a narrow working range, and in a multiplex panel one dilution of the sample rarely suits every protein. Tighe et al. 2015 call a single dilution factor for a whole panel often unfeasible. The large platforms work around it. Olink Explore splits each panel into blocks of proteins with similar concentrations and runs each block at its own dilution, and the SomaScan 11K assay measures each aptamer in one of three serial dilutions.
Where each approach falls short
Section titled “Where each approach falls short”Affinity assays measure only what is on the panel, and they trust the binder. Geyer et al. 2017 list binding to the wrong protein, interference from background molecules such as triglycerides, and the hook effect as limitations inherent to immunoassays, and add that some protein variants are hard for antibodies to recognize. Tate and Ward 2004 review the interferences in detail, from heterophile antibodies to lipemia.
A genetic variant can change how a binder holds its protein without changing the amount of protein. Eldjarn et al. 2023 call this an epitope effect and found Olink and SomaScan assays similarly exposed to it. The same study compared matched assays on Olink Explore 3072 and SomaScan v4 in plasma from 1,514 people, and the median Spearman correlation across 1,848 proteins was 0.33. Two platforms that both claim to measure a protein can disagree about who has more of it. Katz et al. 2022 found that Olink showed more reliable target specificity, while SomaScan offered better precision and more proteins.
Mass spectrometry has the opposite profile. It is specific and open ended, but in plasma it reaches only the abundant proteins without depletion or fractionation. Geyer et al. 2017 report that targeted MS methods typically do not reach the low ng/ml range without extra sample preparation, while baseline IL-6 sits far below that at about 5 pg/ml. Data dependent acquisition is also semi stochastic, so a peptide seen in one run can be missing from the next, and data independent acquisition was introduced to identify peptides more consistently.
Neither family wins outright. This section starts with affinity assays because many labs run them before they turn to mass spectrometry, and because their analysis problems of curves, limits, plates and censored values come up first. The mass spectrometry lessons follow them.
Next steps
Section titled “Next steps”The next page, Affinity platforms compared, shows how ELISA, Luminex, Olink and SomaScan each turn a binding event into a number.
Further reading
Section titled “Further reading”Geyer and colleagues, 2017, review plasma proteomics, its concentration range, and approaches to protein quantification. Their comparison of abundant plasma proteins with low-concentration cytokines explains the measurement problem introduced here. Read their discussion of internal standards for the distinction between relative abundance and an absolute concentration; the instrument performance they report belongs to 2017.